T1 - Comparison of retinal thickness measurements among four different optical coherence tomography devices
AU - Nam, Ki Tae
AU - Yun, Cheolmin
AU - Seo, Myungho
AU - Ahn, Somin
AU - Oh, Jaeryung
N1 - Publisher Copyright:
© The Author(s) 2024.
PY - 2024/12
Y1 - 2024/12
N2 - We sought to compare the retinal thickness measurements collected using different optical coherence tomography (OCT) devices. This prospective study included 21 healthy cases, and the retinal thickness was measured using the PLEX Elite (Carl Zeiss Meditec, Dublin, California, USA), DRI OCT-1 Atlantis (Topcon Corp, Tokyo, Japan), Cirrus 5000 HD-OCT (Carl Zeiss Meditec), and Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany), respectively. The mean central retinal thickness (CRT) and mean retinal thickness of the Early Treatment of Diabetic Retinopathy Study (ETDRS) area were compared. The CRT varied significantly among the different OCT devices (P < 0.001). Post-hoc analysis revealed that the CRT measured using PLEX Elite (278.95 ± 20.04 µm) and Spectralis (271.86 ± 17.92 µm) were similar, and both were greater than the CRT measurements of DRI OCT-1 (239.57 ± 21.06 µm) and Cirrus (256.76 ± 17.82 µm). Additionally, the mean retinal thickness in each ETDRS area showed significant differences among the four devices (all P < 0.001). The mean retinal thickness measured varied according to the device used, and this needs to be considered when comparing retinal thickness measurements taken with different devices.
AB - We sought to compare the retinal thickness measurements collected using different optical coherence tomography (OCT) devices. This prospective study included 21 healthy cases, and the retinal thickness was measured using the PLEX Elite (Carl Zeiss Meditec, Dublin, California, USA), DRI OCT-1 Atlantis (Topcon Corp, Tokyo, Japan), Cirrus 5000 HD-OCT (Carl Zeiss Meditec), and Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany), respectively. The mean central retinal thickness (CRT) and mean retinal thickness of the Early Treatment of Diabetic Retinopathy Study (ETDRS) area were compared. The CRT varied significantly among the different OCT devices (P < 0.001). Post-hoc analysis revealed that the CRT measured using PLEX Elite (278.95 ± 20.04 µm) and Spectralis (271.86 ± 17.92 µm) were similar, and both were greater than the CRT measurements of DRI OCT-1 (239.57 ± 21.06 µm) and Cirrus (256.76 ± 17.82 µm). Additionally, the mean retinal thickness in each ETDRS area showed significant differences among the four devices (all P < 0.001). The mean retinal thickness measured varied according to the device used, and this needs to be considered when comparing retinal thickness measurements taken with different devices.
UR - http://www.scopus.com/inward/record.url?scp=85185131519&partnerID=8YFLogxK
U2 - 10.1038/s41598-024-54109-6
DO - 10.1038/s41598-024-54109-6
M3 - Article
C2 - 38347154
AN - SCOPUS:85185131519
SN - 2045-2322
VL - 14
JO - Scientific Reports
JF - Scientific Reports
IS - 1
M1 - 3560
ER -