Abstract
We sought to compare the retinal thickness measurements collected using different optical coherence tomography (OCT) devices. This prospective study included 21 healthy cases, and the retinal thickness was measured using the PLEX Elite (Carl Zeiss Meditec, Dublin, California, USA), DRI OCT-1 Atlantis (Topcon Corp, Tokyo, Japan), Cirrus 5000 HD-OCT (Carl Zeiss Meditec), and Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany), respectively. The mean central retinal thickness (CRT) and mean retinal thickness of the Early Treatment of Diabetic Retinopathy Study (ETDRS) area were compared. The CRT varied significantly among the different OCT devices (P < 0.001). Post-hoc analysis revealed that the CRT measured using PLEX Elite (278.95 ± 20.04 µm) and Spectralis (271.86 ± 17.92 µm) were similar, and both were greater than the CRT measurements of DRI OCT-1 (239.57 ± 21.06 µm) and Cirrus (256.76 ± 17.82 µm). Additionally, the mean retinal thickness in each ETDRS area showed significant differences among the four devices (all P < 0.001). The mean retinal thickness measured varied according to the device used, and this needs to be considered when comparing retinal thickness measurements taken with different devices.
| Original language | English |
|---|---|
| Article number | 3560 |
| Journal | Scientific Reports |
| Volume | 14 |
| Issue number | 1 |
| DOIs | |
| State | Published - Dec 2024 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Fingerprint
Dive into the research topics of 'Comparison of retinal thickness measurements among four different optical coherence tomography devices'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver