Design of pullback device in intravascular ultrasound system using longitudinal length variation mechanism for rotating axis

Hong Seok Lim, Sung Min Kim

Research output: Contribution to journalArticlepeer-review

Abstract

The longitudinal length variation for rotating axis can be found in pullback device of the Intravascular ultrasound system (IVUS) which is a medical device. For the design of pullback device in intravascular ultrasound system, we have developed a mechanism which is possible the longitudinal length variation in the axis of high-speed rotation as more than 30 times per second. The mechanism prevents the folding and bending of the signal connection line employing helix structures in longitudinal length variation. In the study, We present the longitudinal length variation mechanism for rotating shaft as be able to improve the disadvantage of the existing commercial pullback devices and is to design a pullback device using this mechanism. Also, using the helix structure, this study proposes the mechanism to minimize bending of rotating axis that can occur when the longitudinal length of the axis of rotation is varied.

Original languageEnglish
Pages (from-to)1569-1573
Number of pages5
JournalInternational Journal of Precision Engineering and Manufacturing
Volume17
Issue number11
DOIs
StatePublished - 1 Nov 2016

Keywords

  • Helix structure
  • Intravascular ultrasound
  • Longitudinal length variation
  • Pullback device

Fingerprint

Dive into the research topics of 'Design of pullback device in intravascular ultrasound system using longitudinal length variation mechanism for rotating axis'. Together they form a unique fingerprint.