Skip to main navigation Skip to search Skip to main content

Refractive prediction of four different intraocular lens calculation formulas compared between new swept source optical coherence tomography and partial coherence interferometry

  • Kim's Eye Hospital

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

Purpose To compare the biometry and prediction of postoperative refractive outcomes of four different formulae (Haigis, SRK/T, Holladay1, Barrett Universal II) obtained by swept-source optical coherence tomography (SS-OCT) biometers and partial coherence interferometry (PCI; IOLMaster ver 5.4). Methods We compared the biometric values of SS-OCT (ANTERION, Heidelberg Engineering Inc., Heidelberg, Germany) and PCI (IOLMaster, Carl Zeiss Meditec, Jena, Germany). Predictive errors calculated using four different formulae (Haigis, SRKT, Holladay1, Barrett Universal II) were compared at 1 month after cataract surgery. Results The mean preoperative axial length (AL) showed no statistically significant difference between SS-OCT and PCI (SS-OCT: 23.78 ± 0.12 mm and PCI: 23.77 ± 0.12 mm). The mean anterior chamber depth (ACD) was 3.30 ± 0.04 mm for SS-OCT and 3.23 ± 0.04 mm for PCI, which was significantly different between the two techniques. The mean corneal curvature also differed significantly between the two techniques. The difference in mean arithmetic prediction error was significant in the Haigis, SRKT, and Holladay1 formulae. The difference in mean absolute prediction error was significant in all four formulae. Conclusions SS-OCT and PCI demonstrated good agreement on biometric measurements; however, there were significant differences in some biometric values. These differences in some ocular biometrics can cause a difference in refractive error after cataract surgery. New type SS-OCT was not superior to the IOL power prediction calculated by PCI.

Original languageEnglish
Article numbere0251152
JournalPLoS ONE
Volume16
Issue number5 May
DOIs
StatePublished - May 2021

Fingerprint

Dive into the research topics of 'Refractive prediction of four different intraocular lens calculation formulas compared between new swept source optical coherence tomography and partial coherence interferometry'. Together they form a unique fingerprint.

Cite this