Treatment of femoral shaft fractures with a titanium intramedullary nail

Gun Il Im, Seong Ryong Shin

Research output: Contribution to journalArticlepeer-review

23 Scopus citations

Abstract

Ninety-nine femoral shaft fractures were treated with locked intramedullary nails made from titanium alloy. One of the distal interlocking screws failed in six fractures (6%) and both screws failed in two fractures (2%). Delayed union was associated with all of the eight fractures that had locking screw failure. Young, heavier patients who had nails of small diameter had an increased risk of screw failure. Additional surgery was needed when both screws failed. The authors still use this nail, but currently prefer to ream the medullary canal more so that larger nails can be inserted. Decisions concerning weightbearing are made on an individual basis for each patient, and currently full weightbearing is delayed for young, active, and heavy patients. Two distal interlocking screws should be inserted for treatment of femoral shaft fracture when a Ti locked intramedullary nail is used.

Original languageEnglish
Pages (from-to)223-229
Number of pages7
JournalClinical Orthopaedics and Related Research
Volume401
DOIs
StatePublished - 2002

Fingerprint

Dive into the research topics of 'Treatment of femoral shaft fractures with a titanium intramedullary nail'. Together they form a unique fingerprint.

Cite this